ExcerptAims: To compare a Western Australian made Springfusor 30 mechanical pump to a Graseby MS16A syringe driver.
Each device was assessed for:
1) Reliability
2) Ease of use by professionals
3) Degree of patient satisfaction
The real cost of each delivery system was also determined.
Method: Fifty patients were included in this study. Patients commencing a subcutaneous infusion were randomised to use one of the pumps. After three days the other delivery system was substituted. The volume of fluid infused by each pump was measured every two hours. Each day nursing staff rated their satisfaction with the pump preparation and drug administration. At the end of the trial the patient decided which pump they preferred.
Results: The Springfusor was as reliable as the Graseby in delivering the prescribed daily dose of medication. Although the nursing staff were satisfied with the Springfusor, they regularly rated the Graseby as more acceptable. The patients were comfortable with either device.
Conclusions: The Springfusor is a clinically effective and economic alternative method of subcutaneous drug administration. It is acceptable to the health professionals operating it, and the patients using it.